Let's Chat

Any Query ?

Ion Chromatography NIC-100

Ion Chromatography NIC-100

Ion Chromatography NIC-100

Ion Chromatography NIC-100 works on a pump flow rate of 0.001 to 12.000 mL/min. Offers conductivity cell detection range of 0 to 45000μS/cm. Offers options for anionic and cationic electrolytic self-regeneration suppressors. The column oven ensures consistent eluent and column temperatures. Our ion chromatography flow paths feature standard PEEK and PTFE materials. Our system combines a pump, six-way valve, column, suppressor, and detector.

Specifications
Qualitative Repeatability ≤ 1%
Quantitative Repeatability ≤ 1%
Pump Flow Rate 0.001 to 12.000 mL/min
Pump Head Pressure 0 to 40 MPa
Flow Setting Error < 0.2%
Flow Stability < 0.2%
Pressure Accuracy 0.01Mpa
Pressure Pulsation < 0.25%
Detector Baseline Noise ≤ 0.0005μS/cm
Detector Baseline Drift ≤ 0.005μS/cm per 30min
Detector Resolution 0.00238nS/cm
Minimum Detectable Concentration (Cl): ≤ 0.0005ug/mL
(Li): ≤ 0.0002ug/mL
(NO₃⁻): ≤ 0.001μg/mL
(Na⁺): ≤ 0.001μg/mL
(Ca₂⁺): ≤ 0.005μg/mL
Conductivity Cell Volume ≤ 0.4μL
Conductivity Cell Detection Range 0 to 45000μS/cm
Conductivity Cell Temperature Error ±0.01℃/hr
Conductivity Cell Temperature Resolution 0.01℃
Conductivity Cell Temperature Range 5℃ to 65 ℃
Conductivity Cell Compensation Range 0 to 3.0%
Conductance Cell Stability 0.05℃
Conductivity Cell Voltage 15M
Suppressor Current Source Range 0 to 500mA, increment 0. 1mA
Suppressor Dead Volume 50uL
Suppressor Inhibition Capacity 170ueq/min
Suppressor Withstand Pressure 3MPa
Column Oven Temperature Range RT +5℃ to 70℃;
Column Oven Temperature Stability ≤ 0. 1℃
Column Oven Temperature Error 0.05℃
Column Oven Compatibility 250mm and 150mm columns
Ambient Temperature 10 to 40℃
Relative Humidity 20 to 85%
Power Supply 220V, 50HZ
Mainframe Dimension 1390 × 590 × 432 mm
Pump Dimension 212 × 570 × 430 mm
ECG Dimension 212 × 570 × 430 mm
Detector Dimension 442 × 425 × 590 mm
Net Weight 115 Kg
Gross Weight 195Kg
Features
Applications

Ion Chromatography is used for the detection and analysis of anions (e.g. F⁻, Cl⁻ , Br⁻ , BrO3⁻ , NO3⁻, NO2⁻ , SO42⁻ , PO43⁻ , disinfection by-products, etc.) and cations (e.g. Li⁺ , Na⁺ , NH4⁺ , K⁺ , Mg2⁺ , Ca2⁺ , etc.) in industries like environmental, pharmaceutical, and food samples.

Standard Accessories

Pump Unit × 1 piece


Liquid Pump and mixer × 2 pieces
Eluent generator × 2 pieces
Autosampler (120 vials, sample bottle: 2mL) × 1 piece
Detection and chromatographic Mainframe × 1 piece
Conductivity detector × 2 pieces
Anion Suppressor × 2 pieces
Cation Suppressor × 2 pieces
guard column × 1 piece
Install reagent pack × 1 piece