WhatsApp
Gas Chromatography Mass spectrometry NGMS-100

Gas Chromatography Mass spectrometry NGMS-100

Gas Chromatography Mass spectrometry NGMS-100 features a maximum inlet temperature of 450°C for efficient sample vaporization and accurate results. It is equipped with an electronic pressure flow control system for precise regulation and consistent flow rates. Our unit includes a high-energy dynode electron multiplier for improved sensitivity and detection of trace analytes. It has full-scale gauges for real-time monitoring of vacuum levels.


$56,875.00

Catalog

Specifications

GC Inlet Temperature Maximum 450°C
GC Pressure Range 0 to 100 psi
± 0.002psi
GC Pressure Control Mode Electronic Pressure Control (EPC), Support CV and CC
GC Split Mode Split/Splitless, Max. Split Ratio, 1000:1
GC Column Oven Working Temperature Room Temperature 4°C to 450 °C
GC Heating Rate Up to 120 °C/min
GC Temperature Programing 7 Stages/8 Platforms
MS EI Source Ionization Energy 5 W to 250 W, V (Adjustable)
MS Mass Range 1.5 to 1000 amu
MS Resolution Unit Resolution(Full width at half maximum)
MS Ion Source Temperature 100 to 3500 °C
MS Filament Emission Current 0 to 350 pA
GC-MS Interface Temperature Maximum 450 °C
MS Stability ± 0.10 amu/48 hrs
MS Sensitivity Full scan, 1 pg OFN at m/z 272 with S/N ≥30:1 (RMS)
MS Scan rate Up to l000 amu/s
MS Vacuum Turbo molecular pump(67L/s)
MS Detector High energy dynode electron multiplier
Power 220V
Dimension 720 × 620 × 820 mm
Net Weight 110 kg
Gross Weight 160 kg

Features

  • EI filament for efficient electron emission
  • RF power supply digital compensation
  • High-efficiency ionization source
  • Chem Analyst software for data analysis
  • Vacuum system with mechanical and turbo pumps
  • Self-protection system
  • Mass analyzer with a pre-filter

Applications

  • Gas Chromatography Mass Spectrometry identifies and quantifies chemical compounds by separating and analyzing them. It is used in environmental analysis, pharmaceuticals, and forensic science.

Standard Accessories

Mechanical Pump × 1 pc GC-MS Column × 1 pc Network Cable × 3 pcs Series Port Cable × 1 pc Power Cord × 2 pcs Connection Cord × 1 pc Sylphon Bellows (1 Meter) Clamping Rings × 2 pcs Seal × 2 pcs Wrench × 3 pcs Tweezers × 1 pc Cross Screwdriver × 2 pcs Straight Screwdriver × 1 pc Monkey Wrench × 1 pc Allen Wrench × 1 pc Sleeve × 1 pc 15% Graphite /85% Vespel × 10 pcs Syringe × 1 pc Column Cutter × 1 pc Liner × 1 pc Fuse × 2 pcs Sylphon Bellows (1.5 Meter) Sylphon Bellows Connector × 1 pc Hose clamps × 1 pc Ion Source Holder × 1 pc Column Mounting Jig × 1 pc Stainless Steel Tube (7 Meter) 1/8"Ferrule1/8” × 1 pc Valve Connector × 2 pcs O-ringO × 11 pcs Glass Wool × 1 pc Front Ferrule × 3 pcs Back Ferrule × 3 pcs Injection port Wrench × 1 pc Glass Liner × 1 pc Moisture Trap × 1 pc Oxygen Trap × 1 pc Trap Bracket × 1 pc Ties × 8 pcs Mechanical pump outlet joint × 2 pcs Adapter × 1 pc Ethernet switch × 1 pc Injection Septum × 1 pc

Optional Accessories

Auto Sampler

Ask Expert

Labnics © 2025. All Rights Reserved.