Any Query ?
Inductively Coupled Plasma Emission Spectrometer NICP-101 provides a detection limit between 1 ppb and 10 ppb, ensuring accurate analysis of trace elements. It features a spectral line range of 165 nm to 950 nm for comprehensive evaluation. The light chamber includes an independent gas path for filling with nitrogen or argon for optimal performance. Our unit incorporates a refined optical design with aspherical components to enhance imaging quality.
Element Detection Limit | 1ppb to 10ppb for most elements |
Repeatability | RSD < 1% |
Stability | RSD < 1.5% over 2 hours |
Solid Content Measurement Range | 0.001% to 70% |
Liquid Content Measurement Range | 0.01ppm to several thousand ppm |
Operating temperature | 15℃ to 30℃ |
Operating humidity | ≤ 70% |
Atmospheric pressure | 86 to 106 kPa |
Relative humidity during storage and transportation | ≤ 85% |
Storage and transportation temperature | 5℃ to 40℃ |
Optical System | Medium-step two-dimensional spectroscopic optical system Multi-point filling technology of the light chamber Radial and axial observation Spectral line range: 165nm to 950nm Focal length: 400mm Temperature control: 38±0.1℃ (Adjustable) |
Stray light | Equivalent background concentration of 10000μg/ml Ca solution at As189.042nm is < 2 μg/ml |
Prism | Ultra-pure SiO2 material |
Grating | Echelle grating Size: 100mm × 50mm |
Detector | Large-area CCD Number of pixels: 1024 ×1024 Pixel area: 24µm × 24µm Operating temperature: < -40℃ Stabilization time: < 3 minutes Quantization efficiency: >75% Anti-saturation overflow |
RF Generator | All-solid-state, automatic matching Power output: 700W-1600W, continuously adjustable by 1W Power stability: ≤0.1% Oscillation frequency: 27.12MHz Frequency stability: ≤0.01% Electromagnetic field leakage radiation intensity: ˂0.5 V/m |
Sample Introduction System | Optimized for organic solvents, high-salt/complex matrix, and HF-containing samples Integrated torch, easy to maintain, three concentric quartz torch with 20mm outer diameter Torch Coil: 3 turns Atomizer: concentric atomizer or parallel channel atomizer, outer diameter 6mm, optional standard atomizer, high salt atomizer, hydrofluoric acid atomizer |
Sample Introduction System Components | Peristaltic pump: 4 channels, 12 rollers, continuously adjustable speed Spray chamber: cyclonic spray chamber, optional double-cylinder spray chamber, and HF-resistant spray chamber |
Argon Gas Consumption | 8L/min to 18L/min |
Cooling Gas Flow Rate | 0.00L/min to 20.00L/min, accuracy 0.01L/min |
Auxiliary Gas Flow Rate | 0.00L/min to 2.00L/min, accuracy 0.01L/min |
Carrier Gas Flow Rate | 0.00L/min to 2.00L/min, accuracy 0.01L/min |
Waterway System | Two-inlet, six-outlet integrated feedback diverter system |
Control System | Multi-window software with over 70,000 spectral lines Analysis methods: Standard curve and standard addition method Data processing methods: Indoor blank, internal standard and interference correction methods |
Certifications | ISO 9001, ISO 14001, ISO 45001, FCC, EU CE, EU RoHS, US FDA |
Power supply | 220±22VAC 50±1Hz |
Dimensions (L × W × H) | 1270 × 720 × 850 mm |
Packaging Dimensions (L × W × H) | 1500 × 900 × 1060 mm |
Net Weight | 192 kg |
Gross Weight | 250 kg |
Labnics product catalog
Generated on: 12/02/2025
Labnics Ltd.
Unit 2D Station House, 1 Pembroke Broadway, Camberley,
Surrey GU15 3XD United Kingdom