Dry Block Temperature Calibrator NDTC-101 offers a temperature range from RT+10â to 700â for reliable calibration. It features dual-zone temperature control for uniform heating and consistent results. This calibrator supports programmable multi-point calibration for thermocouples and resistance sensors. It utilizes a platinum resistance sensor for accurate temperature measurements. Our unit is equipped with a 4.0-inch touchscreen for intuitive and efficient operation.
Temperature Control Range | RT+10℃ to 700℃ |
Operational Temperature | 0 to 50 ℃ |
Display Resolution | 0.001℃ |
Built-In Reference Temperature Control Accuracy | ± 0.15℃ at 100℃ ± 0.60℃ at 600℃ |
External Reference Temperature Control Accuracy | ± 0.10℃ at 100℃ ± 0.20℃ at 600℃ |
Temperature Fluctuation | ± 0.03℃ at 100℃ ± 0.07℃ at 600℃ |
Temperature Difference Between Holes | 0.03℃ at 100℃ 0.10℃ at 600℃ |
Heating Time | 23℃ to 700℃: 25min |
Cooling Time | 700℃ to 100℃: 35min |
Stabilization Time | ≤ 10min |
Temperature Block Diameter/Insertion Depth | φ26mm/155mm |
Default Display | 4.0-inch industrial touch screen, Resolution 800 × 480 pixel |
Communication Method | RS485 |
Power Consumption | 1200W |
Power Supply | 220VAC±10%, 50Hz |
Dimensions (H × W × D) | 360 × 180 × 270 mm |
Weight | 9.2 kg |
Labnics product catalog
Generated on: 14/01/2025
Labnics Ltd.
Unit 2D Station House, 1 Pembroke Broadway, Camberley,
Surrey GU15 3XD United Kingdom
Email: info@labnics.com | Website: www.labnics.com | Phone: +44 755 143 4216