Any Query ?
Airborne Particle Counter NAPC-400
Airborne Particle Counter NAPC-400 is used to measure dust particles inside a clean room environment. Its laser particle counter uses a semiconductor laser with a light source and a large touch-screen display. It is small in size and with light weight, highly precise and easy to operate. The microprocessor can control, store and print the measurement results, the test environment is very clean convenient. In addition, special software can be downloaded via computer to store the data. A high volt DC power source is applied which results in maximum consumption that is around 80 W.
Flow Rate | 50 L / min import sampling pump |
Sample Time | Between 6 to 999999 s optional (Selectable) |
Alarm | Class |
Operating Time | 5 hours |
Display | LCD Display |
Weight | 7.5 kg |
Outside Dim | 320×272×240 mm |
Power Source | DC 16.8 V (Built in Lithium Battery) |
Max. Consumption | 80 W |
Size Channels | 0.5, 1, 3, 5, 10, 25 µm |
Calculation | 95 % UCL Calculation of Particle Concentration can be Displayed Directly |
Zero Count | ≤ 10 min |
Location Labels | 0 to 9999 |
Printer | Built –in Thermal Printer |
Data Storage | 20000 Group / section 300 UCL (with a PC dedicated communication software |
Output | USB |
Light Source | Laser Diode (30000 hours mean time between failures) |
Maximum Rated Particle Concentration | 35000 particles / L |
Ambient Temperature | 10 to 35 °C |
Humidity | 20 to 75 % RH |
Atmospheric Pressure | 86 to 106 kPa |
Airborne Particle Counter is widely used in various fields such as medicine, optical, chemical, food, cosmetics, e-health, biological products, aerospace and other sectors.
Particle Channel | 0.3,2.5,10 μm |
Mass Concentration | PM2.5: 0~500 ug/m3; PM10: 0~500 ug/m3 |
Flow Rate | 0.1 ft³ (2.83 L/min) |
Deviation | 5 % ; 2,000,000 particles per ft³ |
Particle Channel | 2.5,10 μm |
Mass Concentration | PM2.5: 0~500 ug/m3; PM10: 0~500 ug/m3 |
Resolution | 1 ug/m3 |
Temperature Range | 0 °C - 50 ℃ |
Light source | Laser diode (lifetime greater than 35000 hours) |
Particle channel | 0.3, 0.5, 1, 3, 5, 10 µm |
Sampling period | 1 to 10000 (s) time-delay: 0 to 99 (s) Self-cleaning time: 10 min |
Sampling flow | 2.83 L / min (0.1 cfm) Imported pump |
Particle Channel | 0.3, 0.5, 1.0, 2.5, 5.0, 10 μm |
Mass Concentration | PM2.5: 0~500 ug/m3; PM10: 0~500 ug/m3 |
Flow Rate | 0.1 ft³ (2.83 L/min) |
Deviation | 5 % ; 2,000,000 particles per ft³ |
Test range | 0.1 μm to 800 μm |
Number of detector channels | 70 |
Accuracy error | 1% (National standard sample D50 value) |
Repeatability error | 0.5% (National standard sample D50 value) |
Labnics Liquid Particle Counter are designed for the precise detection and analysis of particles in liquids across a broad size range. Our units are suitable for analyzing multiple samples simultaneously for enhanced performance. These Liquid Particle Counters provide automatic data storage and retrieval for efficient management of information. They include color touchscreens, allowing for easy operation and clear data visualization.
Particle size range | 0.8 to 600 µm (calibrated according to 1-100um or 4-70um(c)) |
Sensitivity | 0.1μm |
Detection channels | 8 to 64 channels are optional, and the particle size can be set arbitrarily |
Sampling volume | 0.2 to 1000ml<br/> Interval 0.1mL |
Particle Size Range | 0.2 to 2 μm |
Detection Channels | 16 |
Maximum Concentration | 10000 particles/ml |
Accuracy | ± 10% |
Labnics Portable Oil Particle Counter boasts a wide measurement range, capable of detecting particles making it versatile for different oil types. Our particle counters are integrated with optical or laser sensors to ensure reliable and repeatable measurements. They are designed for efficient and accurate monitoring of oil contamination levels in various industries. These are equipped with a user-friendly touchscreen interface, enabling quick navigation and real-time monitoring. Features built-in data logging and storage capabilities allowing users to track contamination levels over time and generate detailed reports. These are crafted with durable materials and withstand harsh environments, maintaining reliable performance.
Particle Size Range | 0.8 to 600 µm |
Detection Channel | 8 channels |
Sampling Volume | 0.2 to 1000 mL |
Sampling Accuracy | Better than ± 1% |