| Sample capacity |
16 wells (0.2mL PCR tube or strips) |
| Reaction volume |
10 to 50μl |
| Thermal cycling technique |
Peltier |
| Cooling method |
Liquid cycle refrigeration |
| Maximum ramp rate |
8℃/s |
| Temperature accuracy |
±0.2℃ |
| Temperature homogeneity |
±0.2℃ at 60°C ±0.2℃ at 95°C |
| Light source |
2 or 4 monochrome efficient LEDs |
| Optical detector |
High-performance MPPC |
| Detection mode |
Ultra-fast time-resolved scanning |
| Fluorescent channel |
FAM/SYBR Green, VIC/JOE/HEX/TET, JUN, ROX/Texas Red (optional), Mustang Purple, Cy5/LIZ (optional) |
| Running time |
<30 minutes (combined with rapid kit) |
| Sensitivity |
Single copy of the gene, the difference of 1.33 times concentration can be distinguished |
| Dynamic range |
10 orders of magnitude |
| Operation system |
Windows |
| Data analysis modes |
Qualitative Detection, Absolute Quantification, Relative Quantification, Genotyping |
| Noise |
<50 dB |
| Power supply |
220VAC, 50/60hz |
| Dimension |
205×190×98 mm |
FAQ for Real Time PCR NRTP-600
1: How accurate and reliable are the test results?
The Real Time PCR system ensures precise temperature control and delivers consistent performance with a Ct standard deviation of less than 0.2. It also offers single-copy gene detection for high sensitivity and reliability.
2: Is the Real Time PCR NRTP-600 suitable for on-site testing?
Its compact size and lightweight build make it highly portable for use outside traditional laboratories. This makes it ideal for disease control, food safety inspections, and environmental monitoring.
3: How often does the Real Time PCR NRTP-600 require maintenance?
The Real Time PCR system is designed for minimal maintenance with stable long-term performance. Routine checks every few months are sufficient to ensure optimal operation.
4: What parts of the Real Time PCR system need regular cleaning?
The exterior, sample wells, and optical detection areas should be kept clean. Using lint-free wipes and non-corrosive cleaning agents is recommended to avoid damage.
5: How should the Real Time PCR NRTP-600 be stored when not in use?
When not in use, the Real Time PCR system should be stored in a clean, dry environment with stable room temperature. Covering the unit helps protect it from dust and accidental exposure.
6: How accurate and reproducible are the results from this Real Time PCR system?
This system delivers high reproducibility with minimal variation between replicates. Its uniform thermal block design ensures consistent amplification across all wells, which is crucial for sensitive Real Time PCR analysis, including low-copy target detection.
7: How user-friendly is the software interface of this Real Time PCR system?
The software is designed for ease of use, featuring a navigational interface and project management tools. Users can quickly load saved protocols, import sample information, and initiate Real Time PCR runs with minimal input, reducing setup time and human error.
8: Is this Real Time PCR system suitable for point-of-care or field use?
Absolutely. Its lightweight, portable design and battery-powered operation make this Real Time PCR system perfect for point-of-care testing and mobile diagnostic units. It is engineered for stable performance even in remote or non-laboratory settings.
9: What steps should I take if the Real Time PCR system shows inconsistent results?
Ensure that sample volumes are uniform, reagents are fresh, and the instrument is placed on a stable surface. Verifying thermal cycling parameters and proper sealing of reaction tubes can also help improve Real Time PCR consistency.