WhatsApp
Lab Glove Box NGB-200

Lab Glove Box NGB-200

Labnics Lab Glove Box NGB-200 incorporates a 3 mm 304 stainless steel chamber, 0.3 µm filtration, and a Φ377 × 600 mm antechamber for controlled atmosphere processing. This glove box system maintains oxygen and moisture below 0.1 ppm, helping reduce atmospheric exposure, preserve sample integrity, and support handling of oxygen and moisture-sensitive materials.


$26,950.00

Data Sheet

Specifications

H₂O & O₂ Content <0.1 ppm
Leakage Rate <0.001 vol%/h
Thickness 3 mm
Material 304 stainless steel
Antechamber Φ377×600 mm
Filter 0.3 micron
Internal Dimension 1220 × 760 × 900 mm
External Dimension 1300 × 800 × 950 mm

Features

PLC Control System — Automated glove box control

High Precision — Precise operating control

Modular Design — Flexible system configuration

Single-Side Operation — Convenient glove box access

Sturdy Bracket Support — Stable chamber support

Applications

Battery Manufacturing Industries

Materials Science Laboratories

OLED Research Laboratories

Solar Cell Industries

Assembles lithium battery components in a moisture and oxygen-free environment with copper catalyst technology

Processes sensitive materials with ultra-pure controlled atmosphere and touchscreen monitoring

Conducts OLED research with stable inert gas circulation and a removable safety glass window

Develops solar cells in contamination free sealed chamber with automatic pressure regulation

Ask Expert

FAQ for Lab Glove Box NGB-200

1: What design features of the Glove Box improve user safety and ergonomics?

It includes a slanted working surface, removable safety glass front window, and single-side operation, offering comfortable access and safe handling of sensitive substances inside the glove box.

2: What should I do if the Glove Box fails to maintain its internal environment?

If the Glove Box shows unstable atmosphere control, check the seals, verify regeneration cycles, inspect the molecular sieve and copper catalyst, and ensure the vacuum pump is functioning properly.

3: Can the Glove Box NGB-200 operate with different inert gases?

Yes, the Glove Box can operate with inert gases like nitrogen or argon, depending on your application requirements and compatibility with installed purification units.

4: What materials are used for the Glove Box construction?

The Glove Box is typically constructed from stainless steel or coated steel with durable seals and safety glass, ensuring chemical resistance and long operational life.

5: What kind of maintenance does the Glove Box Glove Box NGB-200 require?

Routine maintenance includes filter inspection, leak checks, regeneration of the purification system, and cleaning of the interior with compatible solvents.

Related Models Of Lab Glove Box

Labnics Laboratory Glove Box utilizes closed-loop gas purification, vacuum transfer airlocks, and automatic pressure regulation to maintain oxygen and moisture levels of 0.1 ppm or lower within the work chamber. Available in acrylic and 304 stainless steel configurations. These glove box systems feature 0.3 µm HEPA filtration, a leakage rate of 0.001 vol%/h or lower, and inert gas circulation for stable, controlled atmosphere operation.
 

Labnics © 2026. All Rights Reserved.