Labnics Lab Glove Box NGB-200 incorporates a 3 mm 304 stainless steel chamber, 0.3 µm filtration, and a Φ377 × 600 mm antechamber for controlled atmosphere processing. This glove box system maintains oxygen and moisture below 0.1 ppm, helping reduce atmospheric exposure, preserve sample integrity, and support handling of oxygen and moisture-sensitive materials.
FAQ for Lab Glove Box NGB-200
1: What design features of the Glove Box improve user safety and ergonomics?
It includes a slanted working surface, removable safety glass front window, and single-side operation, offering comfortable access and safe handling of sensitive substances inside the glove box.
2: What should I do if the Glove Box fails to maintain its internal environment?
If the Glove Box shows unstable atmosphere control, check the seals, verify regeneration cycles, inspect the molecular sieve and copper catalyst, and ensure the vacuum pump is functioning properly.
3: Can the Glove Box NGB-200 operate with different inert gases?
Yes, the Glove Box can operate with inert gases like nitrogen or argon, depending on your application requirements and compatibility with installed purification units.
4: What materials are used for the Glove Box construction?
The Glove Box is typically constructed from stainless steel or coated steel with durable seals and safety glass, ensuring chemical resistance and long operational life.
5: What kind of maintenance does the Glove Box Glove Box NGB-200 require?
Routine maintenance includes filter inspection, leak checks, regeneration of the purification system, and cleaning of the interior with compatible solvents.