Laboratory Equipment
Vertical Autoclave NVA-400 is designed with a 60 L capacity and functions within an environmental temperature range of 5 to 40°C. It has a mirror-polished inner liner for a clean chamber. This autoclave includes an HD colour LCD touchscreen for real-time parameter monitoring and multi-level password management for enhanced security. Our unit features a flap-type door with mechanical and electronic interlocking for safety.
| Capacity | 60 L |
|---|---|
| Environment Temperature | 5 to 40°C |
| Environment Humidity | 20 to 80%RH |
| Sterilization Temperature | 105 to 135°C |
| Temperature Fluctuation | Less than ±1°C |
| Temperature Uniformity | Less than or equal to ±1°C |
| Melting Temperature | 50 to 100°C |
| Heat Preservation Temperature | 40 to 60°C |
| Sterilization Timer Range | 0 to 240 minutes |
| Melting Timer Range | 0 to 600 minutes |
| Heat Preservation Timer Range | 0 to 600 minutes |
| Reservation Range | 99h59 min |
| Design Pressure | 0.28 MPa |
| Working Pressure | 0.212 MPa |
| Cavity Size | Ø380 × 555 mm |
| Number of Baskets | 2 |
| Power Consumption | 3.2 kW |
| Power Supply | 220 ± 10%V, 50 ± 2% Hz |
| Dimensions | 550 × 588 × 830 mm |
| Net weight | 84 kg |
| Gross weight | 108 kg |
Biosafety filter
Filter element (consumables)
Verification lead
Sample probe
Printer
Liquid tray
Explore dependable laboratory equipment and expert support for research, testing and quality-control environments.