Pulse Vacuum Autoclave NPVA-106 offers a 4000L capacity and functions at a working temperature of 139°C. It has a highly efficient air filter with ≤ 2-micron precision, achieving 99.97% sterilization. Our unit features an independent steam inlet design, allowing steam to bypass the jacket. It includes BD test procedures and a vacuum leak test procedure to ensure process integrity. This pulse vacuum autoclave comes with an advanced Siemens PLC and a color touchscreen interface for easy control.
| Capacity | 4000L |
| Pressure of inner chamber | -0.1 to 0.3Mpa |
| Pressure of jacket | 0.3Mpa |
| Designed temperature | 150°C |
| Working temperature | 139°C |
| Working pressure | 0.25Mpa |
| Vacuum leakage rate | ≤0.13KPa/min |
| Ultimate vacuum | -97Kpa |
| Opening pressure for the inner chamber’s safety valve | 0.28Mpa |
| Opening pressure for jacket’s safety valve | 0.28Mpa |
| Precision of sterilization temperature control | 0 to 2°C |
| Temperature uniformity | ±1°C |
| Negative pressure pulsation vacuum amplitude and times | Amplitude -80 to 0Kpa, Times 1 to 99 |
| Times of pulses across pressure | Amplitude -80 to 80Kpa, Times 1 |
| Times of positive pressure pulses | 1 to 3 |
| Steam consumption | 215Kg/C |
| Tap water consumption | 2000Kg/C |
| Power consumption | 380V, 50Hz |
| Power supply | 8kW |
| Internal dimension | 900×1380×3250mm |
| External dimension | 3580×1860×2230mm |
| Packaging dimension | 3550×2000×2200mm |
| Weight | 4500Kg |
Pulse Vacuum Autoclave NPVA-106 is used to sterilize medical, dental, and laboratory instruments, as well as other materials requiring high-level sterilization. It is widely used in healthcare facilities, research laboratories, pharmaceutical companies, veterinary clinics, and food processing industries.