Semi-Automatic Chemistry Analyzer NSCA-200 is capable of measuring sample capacity from 0 to 3000 µL, with a recommended volume of ≥500 µL for optimal performance. It features five standard wavelength filters and three free filter positions for user customization. Our unit uses a halogen lamp to ensure stable and reliable light output for accurate readings. It comes with a 5-inch color LCD touchscreen for easy operation. This analyzer supports both flow cell and cuvette testing modes for versatile analysis.
| Sample capacity | 0 to 3000µL (500 µL recommended) |
|---|---|
| Wavelength | 5 standard filters (340nm, 405nm, 505nm, 560nm, 630 nm) and 3 free positions |
| Wavelength accuracy | ±2 nm, width ≤ 10nm |
| Absorbance range | 0 to 4.500 Abs |
| Absorbance accuracy | outside 0.0001 Abs, inside 0.00001 Ab |
| Carry over | < 1% |
| Repeatability | < 1% |
| Temperature range | 0 to 40°C |
| Incubator temperature | 37°C, 30°C, 25°C |
| Temperature accuracy | ±0.1°C |
| Humidity | ≤ 80% |
| Test mode | Kinetic, End point, Two-point, Absorbance |
| Quality control curve | X̅, SD, CV%, L-J |
| Flow cell | 32 µL quartz glass, 10 mm |
| Heating source | Philips 6V 10W halogen lamp |
| Display | 5 inches color LCD, touchscreen |
| CPU | ARM series high-speed micro CPU |
| Data storage | ≥10,000 |
| Printer | Internal thermal printer |
| Interface | RS232 |
| Power supply | 100 to 240VAC,50 to 60Hz, |
| Overall dimension | 340×270×160mm |
| Weight | 5kg |
| Liver function | ALT, AST, ALP, T-BIL, D-BIL, TP, ALB, TTT, CHE, Nh3 |
| Renal function | BUN, CRE, UA, URE, CO2-CP, γ-G |
| Lipids | T-CHO, TG, HDL-C, LDL-C |
| Cardiac function | CK, CK-MB, LDH |
| Immune proteins | IGA, IGG, IGM, C3 |
| Others | GLU, HGB, AMS |
Explore dependable laboratory equipment and expert support for research, testing and quality-control environments.