Analytical High-performance Liquid Chromatography NHLC-400 Catalog

Analytical High-performance Liquid Chromatography NHLC-400 is an advanced, highly effective analytical HPLC system that includes a UV-Vis detector with a wavelength range of 190 to 700 nm, a DAD (Diode Array Detector) detector with a wavelength range of 190 to 800 nm, and a high-pressure constant flow pump and low-pressure quaternary constant flow pump with a flow range of 0.001 to 10.000 mL/min (step 0.001 mL/min). This HPLC system provides accurate, reliable results with improved working efficiency. Powerful software is integrated to regulate, capture, and analyze data, as well as manage the operation.

FEATURES

  • Comprehensive and Flexible System Configuration
  • Pumping system: high-pressure constant flow pump, low-pressure quaternary constant flow pump
  • Equipped with the isocratic system, binary gradient system, and quaternary gradient system
  • Advanced USB chip and optoelectronic isolation ensure stable communication
  • Powerful software for controlling, capturing, and analyzing data, as well as managing operations
  • Real-time pressure display, along with maximum and minimum pressure warnings, ensures safety
SPECIFICATIONS
High-Pressure Constant Flow Pump
Flow RangeAccuracyPrecisionPressure RangePowerDimension (L × W × H)Weight
0.001 to 10.000 mL/min (step 0.001 mL/min)≤±0.2%≤0.06% RSD0 to 45 MPaAC 220 V±10%, 50 Hz/80 W420 × 300 × 175 mm12 kg
Low-Pressure Quaternary Constant Flow Pump
Flow RangeGradient AccuracyRetention Time RepeatabilityMaximum Working PressureCommunication MethodsMixed VolumePowerDimension (L × W × H)Weight
0.001 to 10.000 mL/min (step 0.001 mL/min)≤±1%≤0.1%63 MpaUDP Network Communication750μLAC 220 V±10%, 50 Hz/100 W400 × 300 × 180 mm 15 kg
UV-Vis Detector
LinearWavelength RangeNoiseDriftCellPowerDimension (L × W × H)Weight
2.0 AU190 to 700 nm≤± 0.5,105 AU≤1.0 × 104 AU/hAnalytical: Optical path 10 mm, 11μL Semi-prep scale: Optical path 2 mm, 6μLAC 220V±10%, 50 Hz/100 W420 × 300 × 175 mm16 kg
Diode Array Detector
LinearWavelength RangeNoiseDriftPowerDimension (L × W × H)Weight
≥2.0 AU190 to 800 nm≤±1.0M05 AU≤2.0 × 104 AU/hAC220 V±10%, 50 Hz/180 W500 × 300 × 175 mm18 kg
Column oven
Temperature RangeAccuracyPrecisionPowerDimension (L × W × H)Weight
Ambient +5°C to 85 °C±0.3 °C≤0.1 °CAC220 V±10%, 50 Hz/130 W420 × 300 × 120 mm7kg
Other parameters
PumpDetectorGradient mixerSolvent bottle traySample injectorColumnWorkstationValve mounting frameSystem kitPowerPackage sizeGross weight
P3100 × 2UV3100 × 1 TD-1-15 × 1 TP3100 × 1Rheodyne 7725i × 1Supersil ODS2 5pm ID4.6 × 200mm × 1W5100 × 1ZJ-1 × 1EClassical 3100 × 1AC220V±10%, 50Hz600 × 440 × 380 mm × 345 kg
APPLICATIONS
It is widely used to separate and analyze a mixture of compounds in multidisciplinary fields like research and development studies, food science, environmental science for analysis and detection of impurities, ligand-exchange chromatography, high-pH anion-exchange chromatography of carbohydrates and oligosaccharides, water purification, pre-concentration of trace components, and ion-exchange chromatography of proteins
OPTIONAL ACCESSORIES
Column oven

Labnics product catalog

Generated on: 12/02/2025

Labnics.com

Labnics Ltd.

Unit 2D Station House, 1 Pembroke Broadway, Camberley,

Surrey GU15 3XD United Kingdom

Email: info@labnics.com | Website: www.labnics.com | Phone: +44 755 143 4216