Any Query ?
Electronic Pipette NELP-101
Electronic Pipette NELP-101 is a single-channel unit that handles volumes ranging from 100 to 1000 μL. This multifunctional unit offers high performance, ensuring accuracy and repeatability in operations. Our pipette is motor-driven, enabling variable aspiration and dispensing speeds for controlled liquid handling. It is equipped with an intuitive menu interface that allows easy adjustment of settings and functions.
No. of channels | 1 |
Volume Range | 100 to 1000 μL |
Increment | 5 μL |
Test Volume | 1000 μL, 100 μL |
Accuracy Error |
1000 μL: ±6.00 μL, ±0.60% 100 μL: ±3.00 μL, ±3.00% |
Precision Error |
1000 μL: ±2.00 μL, ±0.2% 100 μL: ±0.60 μL, ±0.60% |
Electronic pipette is used in laboratories for precise and automated liquid handling. It is widely used in scientific experiments, clinical diagnostics, and pharmaceutical research.
Box
Labnics Multi-Channel Pipettes feature an ergonomic design for comfortable one-handed operation. Their large central control buttons and independent tip ejection enhance usability. They include color-coded tips for easy selection and accuracy. Their durable construction ensures resistance to corrosion and high temperatures. These pipettes can be fully sterilized and provide easy calibration technology. Our Multi-Channel Pipettes deliver precision and reliability for all your laboratory needs.
Volume Range | 100 µl, 500 µl, 1000 µl, |
Minimum Increment | ± 1 µl |
Systematic Error | 100 µl: ± 3.00%<br/> 500 µl: ± 1.00%<br/> 1000 µl: ± 0.60% |
Accuracy | 100 µl: ± 3 µl<br/> 500 µl: ± 5 µl<br/> 1000 µl: ± 6 µl |
No. of channels | 8 |
Volume Range | 0.5 to 10 µL |
Large Increment | 10/5/1 µL |
Accuracy | 1.6/2/8 % |
No. of channels | 8 |
Volume Range | 5 to 50 µL |
Large Increment | 50/25/5 µL |
Accuracy | 0.8/1.4/6 % |
No. of channels | 8 |
Volume Range | 10 to 100 µL |
Large Increment | 100/50/10 µL |
Accuracy | 0.8/0.6/4 % |
No. of channels | 8 |
Volume Range | 20 to 200 µL |
Large Increment | 200/100/20 µL |
Accuracy | 0.8/1.4/4 % |
No. of channels | 8 |
Volume Range | 30 to 300 µL |
Large Increment | 300/150/30 µL |
Accuracy | 0.6/1.2/3 % |
No. of Channels | 12 |
Volume Range | 0.5 to 10 µL |
Large Increment | 10/5/1 µL |
Accuracy | 1.6/2/8 % |
No. of Channels | 12 |
Volume Range | 5 to 50 µL |
Large Increment | 50/25/5 µL |
Accuracy | 0.8/1.4/6 % |
No. of Channels | 12 |
Volume Range | 10 to 100 µL |
Large Increment | 100/50/10 µL |
Accuracy | 0.8/0.6/4 % |
No. of Channels | 12 |
Volume Range | 20 to 200 µL |
Large Increment | 200/100/20 µL |
Accuracy | 0.8/1.4/4 % |
No. of Channels | 12 |
Volume Range | 30 to 300 µL |
Large Increment | 300/150/30 µL |
Accuracy | 0.6/1.2/3 % |
Labnics Pipette Stand efficiently stores and organizes micropipettes, enhancing accessibility. They are compatible with a wide range of pipette series, ensuring versatile use. These units feature a durable construction for long-lasting performance. They include a freely rotating design that allows for easy access, thereby enhancing workflow efficiency. Our pipette holders are tool-free and interchangeable for effortless adaptability for various pipette types.
Design | Rotatable stand |
Capacity | Six single or multi-channel pipettes |
Interchangeability | Tool-free, interchangeable pipette holders |
Compatibility | Holds 6 Eppendorf pipettes |
Labnics Adjustable Volume Pipettes boast a variable volume range and facilitate precise liquid handling. These pipettes allow users to easily select the desired volume, providing flexibility for tasks ranging from micro-sampling to larger liquid transfers. They feature an ergonomic design that ensures comfortable handling, significantly reducing hand fatigue during extended use. Our pipettes have an easy volume adjustment mechanism with a dial or push-button for efficient dispensing and quick settings. Their compatibility with standard pipette tips ensures a secure and leak-proof fit for reliable liquid transfer. Equipped with a digital display, these pipettes clearly show the set volume, enhancing usability and precision. Additionally, they are calibrated to meet international standards, ensuring consistent accuracy in results.
Volume range | 0.5-10µl |
Test volume | 10µl 5µl 1µl |
Micropipette tip | 10µl |
Increment | 0.1µl |
Volume range | 2-20µl |
Test volume | 20µl 10µl 2µl |
Micropipette tip | 200µl |
Increment | 0.5µl |
Volume range | 5-50µl |
Test volume | 50µl 25µl 5µl |
Micropipette tip | 200µl |
Increment | 0.5µl |
Volume range | 10-100µl |
Test volume | 100µl 50µl 10µl |
Micropipette tip | 200µl |
Increment | 1µl |
Volume range | 20-200µl |
Test volume | 200µl 100µl 20µl |
Micropipette tip | 200µl |
Increment | 1µl |
Volume range | 100-1000µl |
Test volume | 1000µl 500µl 100µl |
Micropipette tip | 1000µl |
Increment | 5µl |
Volume range | 1000-5000µl |
Test volume | 5000µl 2500µl 1000µl |
Micropipette tip | 5000µl |
Increment | 50µl |
Volume range | 2-10ml |
Test volume | 10ml 5ml 2ml |
Micropipette tip | 10ml |
Increment | 0.1ml |
Volume Range | 10-100μL | ||||
Increment | 1μL | ||||
Test Volume | 100μL 50μL 10μL | ||||
Maximum permissible systematic error (Inaccuracy) |
|
No. of Channels | 1 |
Volume Range | 0.1 to 2.5 μL |
Increment | 0.05 μL |
Test Volume | 2.5 μL, 1.25 μL, 0.25 μL |
Volume Range | 1000 to 10000 μL |
Increment | 100μL |
Test Volume | 10000μL, 5000 μL, 1000μL |
Systematic Error | ± 60μL, ± 60μL, ± 30μL<br> ±0.60%, ±1.20%, ±3.0% |
Our Fixed Volume Pipette precisely delivers a specific, predetermined volume for accurate results. They feature an ergonomic design that ensures comfortable handling during extended use, reducing hand fatigue and enhancing the user experience. The easy push-button operation allows for efficient and reliable liquid dispensing. Our pipettes are compatible with standard pipette tips, guaranteeing a secure and leak-proof fit for dependable liquid transfer. They are crafted from high-quality, autoclavable materials that are durable and resistant to chemical corrosion, ensuring longevity. Additionally, they are easy to calibrate, low-maintenance, and offer accuracy in liquid handling for various applications.
Volume Range | 10 µl |
Test volume | 10 µl |
Micropipette tip | 10 µl |
Increment | 0.08 µl |
Volume Range | 100 µl |
Test volume | 100 µl |
Micropipette tip | 200 µl |
Inaccuracy | 0.8 μl |
Volume Range | 100µl |
Test volume | 100 µl |
Micropipette tip | 200 µl |
Inaccuracy | 0.5 μl |
Volume Range | 20 µl |
Test volume | 20 µl |
Micropipette tip | 200 µl |
Inaccuracy | 0.12 μl |
Volume Range | 1000 μL |
Test Volume | 1000μL |
Systematic Error | ± 3μL<br> ±0.3% |
Random Error | ±0.2μL<br> ±0.2% |