Laboratory Equipment
Pulse Vacuum Autoclave NPVA-109 offers a 600L chamber and works at a temperature of 139°C. It functions under a high vacuum of -97Kpa with adjustable pressure pulsations. This unit features an independent steam inlet to prevent steam from passing through the jacket. It includes a water-saving system that cuts vacuum pump water consumption by 30%. Our pulse vacuum autoclave comes with a communication port that connects to a quality traceability system for data logging and compliance.
| Capacity | 600L |
|---|---|
| Pressure of inner chamber | -0.1 to 0.3Mpa |
| Pressure of jacket | 0.3Mpa |
| Designed temperature | 150°C |
| Working temperature | 139°C |
| Working pressure | 0.25Mpa |
| Vacuum leakage rate | ≤0.13KPa/min |
| Ultimate vacuum | -97Kpa |
| Opening pressure for the inner chamber’s safety valve | 0.28Mpa |
| Opening pressure for jacket’s safety valve | 0.28Mpa |
| Precision of sterilization temperature control | 0 to 2°C |
| Temperature uniformity | ±1°C |
| Negative pressure pulsation vacuum amplitude and times | Amplitude -80 to 0Kpa, Times 1 to 99 |
| Times of pulses across pressure | Amplitude -80 to 80Kpa, Times 1 |
| Times of positive pressure pulses | 1 to 3 |
| Steam consumption | 35Kg/C |
| Tap water consumption | 320Kg/C |
| Power consumption | 380V, 50Hz |
| Internal dimension | 610×910×1150mm |
| External dimension | 1470×1410×2100mm |
| Packaging dimension | 1820×1520×2180mm |
| Weight | 1250Kg |