Automatic Hematology Analyzer NAHA-202 can analyze 60 samples per hour and comprises 28 parameters with 2 histograms and 2 scattergrams. Generates 3-D Topographic Maps for WBC analysis. Utilizes an Integral Titanium Incubation System. Incorporates Laser Scatter Technology and ensures Constant Linear Sheath Flow. Samples can be sorted by sample number, patient name, type, time frame, etc.
$11,538.00
Specifications
| Sample Volume | 20μL | ||||||||||||||||||||
| Throughput | 60 samples/hour | ||||||||||||||||||||
| Parameters | Regular parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, ROW-SD, ROW-CV, PL T, POW, MPV, PCT, P-LCR, NEU#, NEU%, LYM#, LYM%, MON#, MON%, EOS#, EOS%, BAS#, BAS% Research parameters: ALY#, ALY%, IG#, IG%, along with 2 histograms and 2 scattergrams |
||||||||||||||||||||
| Principle | WBC/RBC/PLT: Electrical impedance method HGB: colorimetric method WBC 5-diff: laser scattering method |
||||||||||||||||||||
| Sample Mode | Whole blood, Pre-diluted blood mode | ||||||||||||||||||||
| Test Mode | CBC+5 different CBC QC and calibration Multiple QC rules such as L-J, X-B |
||||||||||||||||||||
| Technology | Laser Scatter Technology and Constant Linear Sheath Flow | ||||||||||||||||||||
| Performance |
|
||||||||||||||||||||
| Reagents | Cyanide-Free reagents | ||||||||||||||||||||
| Operating Temperature | 15 ℃ to 30 ℃ | ||||||||||||||||||||
| Operating Humidity | 30% to 85% | ||||||||||||||||||||
| Display | External | ||||||||||||||||||||
| Printer | External | ||||||||||||||||||||
| Data storage | 100,000 results | ||||||||||||||||||||
| Data interface | LIS interface | ||||||||||||||||||||
| Power Supply | 110V/220V, 50Hz/60Hz | ||||||||||||||||||||
| Dimensions (L× W × H) | 460 × 570 × 610 mm | ||||||||||||||||||||
| Weight | 53 Kg |
Features
Applications
Ask Expert