Filter Integrity Tester NFIT-101 checks filter integrity within 100 to 10,000 mbar pressure range. It tests bubble point at 100 to 8000 mbar, diffusion flow at 1 to 1000 ml/min, and water intrusion at 0.01 to 100 ml/min. Features an automatic self-test and fault reporting upon startup, maintaining reliability and accuracy. Provides precise offline and online testing with high-accuracy sensors, supporting 100-meter gas path for enhanced flexibility and accuracy. Our tester can test up to 12-core 20-inch filter cartridges, boosting work efficiency.
$15,575.00
Specifications
| Operation Pressure | 100 to 10000 mbar (150psi) |
| Bubble Point Pressure | 100 to 8000 mbar |
| Diffusion Flow | 1 to 1000ml/min |
| Water Intrusion | 0.01 to 100ml/min |
| Net Volume Test Accuracy | ± 4% |
| Bubble Point Test Accuracy | ±50 mbar |
| Diffusion Flow Test Accuracy | ± 4% |
| Water Intrusion Test Accuracy | ±0.01ml |
| Unit | mbar, kPa, psi, kgf/cm² |
| Relative Humidity | 10 to 80% RH |
| Ambient Temperature | +5℃ to 40℃ |
| Function | Manual Bubble Point Test, Basic Bubble Point Test, Extensive Pressure Holding Test, Diffusion Flow Test, Water Intrusion Test,Ultrafiltration Membrane Test |
| Applied Range | Symmetric and asymmetric membrane test, Needle filters, Capsule filters, Flat filters, Cartridge filters (within 12 cores and 20 inches), Ultrafiltration membrane packages, Ultrafiltration columns, Various irregular filters |
| Audit Trial | Audit trail records can exportable and irreversible |
| Reservation Solution(programs) | 1000 sets |
| Audit Trail Record Storage | 5 years |
| Anti-backflow(liquid) Device | Customization |
| Number of Users | 1000 sets |
| Operating System | Linux system |
| Dust and Splash Level | IP54 |
| Print Function | Built-in printer,can control on PC;choose print subject |
| Historical Record | 300000 sets records storage |
| Records Copy | Support USB disk to export (include test curve) |
| Display Screen | High definition 10-inch, color touch screen |
| Serial Connection | RS232/USB/Support wireless connection |
| Work Mode | Online/Offline |
| Applicable Environment | Above D level |
| Power Consumption | 120 W |
| Power | 100 to 240 VAC, 50Hz |
| Dimension | 480 ×300 ×210 mm |
| Weight | 10 Kg |
Features
Applications
Standard Accessories
Integtest Instrument Power cable(s) Inlet pipe Outlet pipe Operation Manual(include maintenance manual) Calibration Certificate Qualified Certificate Printer Paper Touch Pen 3Q template(optional)
Optional Accessories
Printer
Printer paper
Printer cartridge
10" Housing
5" Housing
Ask Expert