Let's Chat

Any Query ?

Generate Catalog For :

Portable Multiparameter Conductivity Meter NPMCM-100

PORTABLE MULTIPARAMETER CONDUCTIVITY METER NPMCM-100

Portable Multiparameter Conductivity Meter NPMCM-100
PORTABLE MULTIPARAMETER CONDUCTIVITY METER NPMCM-100

Portable Multiparameter Conductivity Meter NPMCM-100 is designed with four measurement modes, measures pH, mV, Conductivity and TDS of liquids. Equipped with a large backlit LCD display. Features automatic temperature compensation, manual temperature calibration and stability indicator and reset feature. The unit has USB interface and built-in real-time clock stamps. Offers Power adapter, multi-mode power scheme for battery and computer's USB port that ensures smooth functioning of meter.

Portable Multiparameter Conductivity Meter NPMCM-100
FEATURES
  • Measures pH, mV, Conductivity and TDS of liquids
  • Large backlit LCD display
  • 1 to 5 points push-button calibration with auto-buffer recognition
  • Automatic electrode slope display assists the user in deciding whether to replace the sensor
  • Auto-Hold function freezes stable measured value for easy viewing and recording
  • Manual temperature calibration, selectable temperature unit (°C or °F)
  • Automatic electrode diagnosis
  • Shows help message as an operational guide
  • System menu allows setting the 12 parameters,including the number of calibration points, resolution, stability condition etc.
  • Built-in real-time clock stamps stored data to meet GLP standard, USB interface
  • Multi-mode power scheme
SPECIFICATIONS
pH
RangeAccuracyResolutionCalibration PointsCalibration SolutionsTemperature CompensationCompensation Mode
-2.000~20.000pH±0.002pH0.1, 0.01, 0.001pH, Selectable1 to 5 pointsUSA, NIST, DIN or User-defined0~100°C, 32~212°FManual or Automatic
mV
RangeAccuracyResolution
-1999.9~1999.9mV±0.2mV0.1, 1mV, Selectable
Conductivity
RangeAccuracyCalibration PointsCalibration SolutionsTemperature CompensationTemperature CoefficientCompensation ModesCell ConstantNormalization Temperature
0~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm±0.5% F.S1 to 5 points10µS/cm, 84µS/cm,1413µS/cm,12.88mS/cm,111.8mS/cm0~100°C, 32~212°F, Manual or Automatic0.0~10.0%/°CLinear or Pure WaterK=0.1, 1, 10 or User-defined20°C or 25°C
TDS
RangeAccuracyTDS Factor
0to100ppt (Max. 200ppt, depending on factor setting)±1% F.S0.1~1.0 (Default 0.5)
Temperature
RangeAccuracyResolutionCalibration PointsCalibration Range
0~105°C, 32~221°F±0.5°C, ±0.9°F0.1°C1 pointMeasured value ±10°C
Others
Hold FunctionStability ConditionsCalibration DuePower OffReset FunctionMemoryOutputConnectorPower RequirementsBattery LifeDimensions (L × W × H)Weight
Manual or AutomaticLow or High0 to 31 daysManual or Automatic (10, 20, 30 minutes)YesStores up to 500 data setsUSB Communication InterfaceBNC, 6-pin Mini Plug3×1.5V "AA" BatteriesApproximately 150 hours (Turn off the backlit)170 × 85 × 30 mm300 g
APPLICATIONS

It is mainly used in the detection of pH, mV, Conductivity and TDS of samples from food & beverage industries, water & wastewater treatment plants, pharmaceutical industries, chemicals & petrochemicals, etc.

Labnics product catalog

Generated on: 20/05/2024

Labnics.com

Labnics Ltd.

Unit 2D Station House, 1 Pembroke Broadway, Camberley,

Surrey GU15 3XD United Kingdom

Email: info@labnics.com | Website: www.labnics.com | Phone: +44 755 143 4216