Benchtop Multi-Parameter Meter NBMM-102 measures pH within a range of -2.000 to 20.000 with an accuracy of ±0.002. It offers automatic and manual temperature compensation for accurate results. This meter can store up to 1000 data sets per parameter, adhering to GLP compliance standards. Our unit offers automatic barometric adjustment for precise dissolved oxygen measurements. It features a 7.0-inch high-contrast LCD touchscreen for easy operation.
Test Parameters | pH/EC/ISE/DO/Temp. (mV/ORP/pX/Resistivity/TDS/Sal. /DO Saturation) |
pH | Range: -2.000 to 20.000 pH Resolution: 0.1, 0.01, 0.001 pH Accuracy: ±0.002 pH Calibration Points: Up to 6 Standard Customization: Yes Calibration Reminder: Yes Slope Limit: Yes |
Calibration Solutions | NIST, GB, DIN, USA, MERK, JIS buffers |
mV/ORP | Range: -2000.00 to 2000.00 mV Resolution: 0.1, 0.01 mV Accuracy: reading whichever is greater Calibration Points: 1 custom point (relative mV) EH ORP Mode: Yes |
pX | Range: -2.000 to 20.000 pX Resolution: 0.1, 0.01, 0.001 pX Accuracy: ±0.002 pX Calibration Points: Up to 6 |
ISE | Range: 1.000e-9 to 9.999e+9 Units: mol/L, mmol/L, g/L, mg/L, μg/L, ppm, ppb Resolution: Up to 4 significant digits Accuracy: ±0.3% Calibration Points: Up to 6 |
Conductivity | Range: 0.000 μS/cm to 2000 mS/cm Resolution: 0.001 μS/cm minimum, varies with range selection Accuracy: ±0.5% FS Calibration Points: Up to 5 Settable parameters: Cell constant, reference temperature, temperature compensation coefficient, and TDS factor Standard Recognition: 10 μS/cm, 84 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm; 146.5 μS/cm, 1408 μS/cm, 12.85 mS/cm, 111.3 mS/cm Reference Temperature: 5, 10, 15, 18, 20, 25 ºC Calibration Reminder: Yes |
Resistivity | Range: 5.00 Ω·cm to 100.0 MΩ·cm Resolution: 0.01 Ω·cm minimum Accuracy: ±0.5% FS |
TDS | Range: 0.000 to 1000 ppt Resolution: 0.001 ppm minimum, varies with range selection Accuracy: ±0.5% FS |
Salinity Type | Type: NaCl Range: 0.0 to 80.0 ppt Resolution: 0.1 ppt Accuracy: ±1 ppt |
Dissolved Oxygen | Concentration (Polarographic) Range: 0.00 to 99.99 ppm Resolution: 0.01 ppm Accuracy: ±0.10ppm, 0.00 to 50.00ppm ±0.50ppm, 50.0 to 99.99ppm Calibration Points: Air-saturated water or zero point Auto barometric pressure compensation Selectable pressure unit: kPa, mbar, Torr, Atm. DO Calibration Reminder: Yes |
% Saturation | (Polarographic) Range: 0.0 to 600% Resolution: 0.10% Accuracy: ±2.0% |
Temperature | Range: -10 to 135 ℃, 14 to 275 °F Resolution: 0.1 Accuracy: ±0.1 |
Temperature Compensation | ATC, MTC |
Measurement Reading Mode | Auto Read (Fast, Medium, Slow), Timed, Continuous |
Measurement Reading Prompts | Reading, Stable, Locked |
Data Storage | 1000 Groups |
GLP Features | Yes |
Log Management | Yes |
pH Electrode Input | BNC(Q9) |
DO with Temp. Probe Input | 4-pin aviation connector |
Conductivity with Temp. Probe Input | 5-pin aviation connector |
Outputs | USB: USB 2.0 flash memory device, PC, scanner RS 232: Printer, auto-sampler |
Backlight | Yes |
Auto Shutdown | 1 to 60 min, off |
Date and Time | Yes |
Power | AC Adapter, 100-240V AC input, DC 24V output |
Dimensions | 220 × 195 × 68 mm |
Weight | 950g (2.09 lb.) |