Generate Catalog For :

Laser Particle Counter NLPC-200

PRODUCT CATALOG

Laser Particle Counter NLPC-200

LASER PARTICLE COUNTER NLPC-200

Laser Particle Counter NLPC-200 is a device used to measure the number and size distribution of dust particles per unit volume in clean environment. The instrument adopts the principle of light scattering, when the suspended particles in the air after a photosensitive area, scattering and its particle size with a certain proportion of luminous flux, get treated by photoelectric conversion, amplification and be harvested equivalent diameter and number of particles. It measures particle counts before, during and after the job done and it shows 6 channels size distribution of 0.3, 0.5, 1.0, 2.0, 5.0, 10 m.

FEATURES
  • Distribution error of particle size is 0.5 m 5 m ? 20 %
  • Sampling flow and error 2.83 L / min ? 5 %
  • Maximum limit to measure the particle concentration ? 30 k / L
  • Measurement repeat relative deviation ? 10 %
  • Indication error of particle concentration 0.5 m ? 20 %
  • Range of particle size measured: 0.3, 0.5, 1, 3, 5, 10 m, divided into 6 particle size channel
  • Simple, fast operation with two front-panel buttons
  • Self-cleaning time ? 10min
  • Bluetooth printing (optional) is used to print the data required
  • Colour LCD touchscreen is launched for better enhancement
  • Rubber sampling head: - to insure the service life of the sensor
SPECIFICATIONS
Light source Laser diode (lifetime greater than 35000 hours)
Particle channel 0.3, 0.5, 1, 3, 5, 10 µm
Sampling period 1 to 10000 (s) time-delay: 0 to 99 (s) Self-cleaning time: 10 min
Sampling flow 2.83 L / min (0.1 cfm) Imported pump
Test mode Single, repetitive, continuous, computing, remote
Count mode The cumulative values, the difference and density
Working time 6 hours (Provide a fuel gauge of remaining battery capacity)
Display LCD real time display, a period, the concentration of the real-time display, can display time, date, measurements, temperature and humidity, room number, number of sampling points, sampling battery, state parameters, such as 95 % UCL computing, can directly show the particle concentration m3
Rechargeable battery Lithium ion battery, 7.4 V, 2200 mAh
Ambient Temperature 10 to 40 °C (50 to 104 °F)
Humidity 20 to 90 % RH
Atmospheric pressure 86 to 106 kPa
UCL set Sampling number (A): 2 to 9 point set. Each point of sampling time: (L) 2 to 9 times set Measurement setup: 0 to 999
UCL report forms Meet the standards of ISO 14644-11
Data store Can store 1000 groups data (Circulation buffer) ( Including particle data, environment data, year month day time, sampling number, data port), The data is not lost when power is cut off.
Communication interface RS 230, 9600 Baud rate
Zero count Instrument with level alarm function, can to dean room grade 100, grade 1000, grade 10000, grade 100000, grade 300000, grade 1000000 with excessive alarm and low battery voltage alarm
Superposition errors 5 %, 2,000,000 grain / cubic feet
Print function Cab connect the external printer
Standard configuration Suitcase charger sampling head nul filter RS232 connecting in CD for computer communication software
Sampling position number 1000 groups
Measuring repeated relative deviation ≤ ± 10 %
Weight 1 kg
Overall dimension 220×40×120 mm
Power supply External AC power adapter, AC: 100 V to 245 V, 50/60 Hz to DC: 7.4 V, 1 A
APPLICATIONS

Laser Particle Counter is used in various fields such as electronics, optical, chemical, food, cosmetics, medicine, biological products, aerospace and other departments for public health.

Labnics

Labnics Equipment Ltd.

272 Bath Street, Glasgow, G2 4JR UK

Email: info@labnics.com | Website: www.labnics.com | Phone: +1 403 589 7725