Anaerobic Chamber NAC-100 is anaerobic systems with controlled atmosphere that are also designed with high capacity incubator and large work area. These chambers attain the desired atmospheric conditions by eliminating oxygen down to 0-5ppm by reacting 5% or less hydrogen gas with a palladium catalyst. Equipped with two sets of gloves and an airlock transfer chamber permits the operator to enter the glove box without compromising the atmosphere inside thereby increasing the tactile efficiency. The unit requires minimal daily maintenance and is easy to use for working with oxygen sensitive materials, product containment needs, and/or general isolation control.
Time for creating anaerobic state in sampling chamber | < 5min |
Time for creating anaerobic state in operation chamber | < 1 hour (Oxygen concentration ≤ 1%) |
Time for maintaining anaerobic state in operation chamber | > 12 h (when no supply of mixed gas) |
Temperature sensor | PT100 |
Temperature range | RT ± 3°C - 60°C |
Temperature fluctuation in incubation chamber | ± 0.3°C |
Temperature uniformity in incubation chamber | ± 1°C |
Temperature controller | PID microprocessor controller |
Timer | 1 min - 9999 min |
Window | Toughened glass window |
Operating room material | Cold-rolled steel plate with electrostatic powder paint |
Incubator material | SUS-304 Stainless steel panel |
External material | Cold-rolled steel plate with electrostatic powder paint |
Power supply | AC 110/220V ± 10%, 50/60Hz |
Inner dimension (W× L ×H) | 810 × 650 × 660 mm |
Outer dimension (W× L ×H) | 1210 × 740 × 1410 mm |
Packaging dimensions | 1340 × 880 × 1590 mm |
Net Weight | 320 kg |
Used in clinical or research institutes that deal with cell culture applications that need to accurately control oxygen, carbon dioxide, temperature and humidity.