Class II Biosafety Cabinet NBSC-300

Class II Biosafety Cabinet NBSC-300 is a Class II safety containment cabinet with type B2 safety level with 99.99% efficient HEPA filters. The downflow and inflow air is exhausted through the filters, without any recirculation. As a result, it limits the operator's exposure to even minute quantities of volatile toxic or hazardous chemicals and radio nucleotide by maintaining positive pressure in the workspace. It thereby helps provides suitable work environment to the user and the product against pathogens and potentially hazardous microbial agents or materials.

Specifications

Air exhaust 1
Glass door orientation Sliding door
HEPA filter 99.999 % efficiency at 0.3 µm ( ×2 )
Inflow velocity 0.53 m/s ± 0.025 m/s
Downflow velocity 0.33 m/s ± 0.025 m/s
Exhaust duct 4 meters PC duct
Exhaust duct diameter 300 mm
Illumination lamp LED lamp, 12 W × 2
Illumination ≤ 1000 Lux
UV lamp 30 W
Display LCD
Maximum opening height 420 nm
Window safety height 200 mm
Alarm system Audio – visual alarm
Power consumption 700 W
Power supply AC 220 / 110 V ± 10 %, 50 / 60 Hz ± 1 Hz
Noise level ≤ 58 dB
Cabinet material Cold-rolled steel with anti-bacteria powder coating
Work table material 304 stainless steel
Window material Two-layer laminated toughened glass
Window thickness ≥ 5 mm
Interior dimensions ( W × H × D ) 940 × 600 × 660 mm
Exterior dimensions ( W × H × D ) 1100 × 750 × 2250 mm
Weight 246 kg

Features

  • PID Microprocessor controller
  • Display for exhaust filter and downflow filter pressure, filter life, etc
  • Visual and audio alarm for abnormal airflow velocity, filter replacement, front window, etc
  • 2 500 W waterproof sockets

Applications

Used in microbiology, pharmaceutical, biotechnology, life science and environmental research-based industries in order to safeguard the user, the product and the environment from toxic and hazardous contaminants.