 |
 |
|
| PCR : Real time
|
 |
- Volume adaptation: Software automatically adjusts heat exchange according to differences in sample volumes.
- New multipoint temperature monitoring allows more accurate and uniform temperature control throughout the heating
block.
- Gradient function: Creates a temperature gradient across the sample block.
- Sample preservation: After completion of experiment, the system can automatically refrigerate samples until they are
moved to storage.
- Use of direct current power supply improves thermal efficiency and reduces power consumption by 30%.
- Improved optics: Greater efficiency in fiber optic system provides reduced signal to noise ratio.
- Automatic hot lid: Provides constant pressure and software-managed temperature control.
- Configuration upgradable: Manual version of the instrument can be upgraded to automatic. Systems can be upgraded from
1 wavelength to 4 wavelengths.
- Numerous data ports: USB and Ethernet connections for PC data interface.
- Multiple tube styles the sample block will accept several styles of 0.2ml tubes including 8- tube strips.
- Data protection system preserves experimental data if there is a power interruption during operation.
- Automatic amplitude adjustment the instrument detects the fluorescence strength in the samples and automatically
adjusts to the correct system sensitivity.
- New software new generation software provides even greater convenience and more powerful functions.
- Upload software instructions define experiment instructions with easy-to-use PC-based software, and then upload to
instrument.
- Overheating protection if the operating environment exceeds the temperature range for operation within specification, the
instrument will shut down.
|
|
|
| Technical Specification |
|
|
| |
| MODEL |

|

|
| Sample Capacity |
48×0.2ml |
| Reaction System |
10~100μl |
| Dynamics Range |
10~1010 Copies |
| Max. Channel Number of Fluorescence Detection |
4 Channels |
2 Channels |
| Excitation Light Wave length(450-590mm) |
Standard Positioning:F1:470nm F2:523nm F3:543nm F4:571nm |
Standard Positioning: F1:470nm F2:523nm |
Fluorescence Detecting Wavelength (500-630nm) |
Standard Positioning: F1:525nm F2:564nm F3:584nm F4:612nm |
Standard Positioning: F1:525nm F2:564nm |
| Detecting Fluorescence |
FAM,SYBR Green I,HEX,VIC,TAMRATEX RED,ROX |
FAM,SYBR Green I, HEX |
| Lead Operation |
Automatic |
Manual |
| Temp. Range |
4°~99.9°C |
| Heating Rate (Max) |
≥ 4.0°C/sec |
| Cooling Rate |
≥ 4.0°C/sec |
| Temp. Uniformity of Block |
≤ ±0.3°C |
| Temp. Control Accuracy |
≤ ±0.1°C |
| Gradient Temp. Range |
1°C~24°C |
| Hot-lid Temp. Range |
80°C~120°C |
| Operation System |
Windows 2000/XP |
| Power Supply |
AC220V 50Hz 500W |
| Dimension (L x W x H) mm |
520 × 450 × 320 |
| Net Weight |
25Kg |
| Operating Manual |
 |
| Catalog no. |
45210101 |
45210102 |
| |
|
 |
|
| |
|
|